The Reweighing function allows you to modify your output file if you have had to adjust the
quantity of sample in some of your tubes after the initial gross weighing.
- Determine which tubes need additional processing via analysis of the Gross weight (.GWT)
files. - Add or remove sample in those tubes, and ensure they are returned to their previous
position in the rack. - Select the Start Button, then select both the Gross and Reweigh mode check boxes. (The
reweigh functionality is started in the same manner as Tare and Gross Weigh modes.)
Select the tube(s) within the first rack which need to be reweighed. These are shown as the
black circles. Select OK.
NOTE: The software will only reweigh tubes from one tube rack at a time
4. From the Tray Setup Screen, select the rack’s deck position and type (or bar code scan) the
rack ID. Select OK and the XL20B will begin the reweigh process. In this example, the XL20B
will reweigh the 5 tubes selected within the first rack of tubes selected during the Tray Setup
process.
5. Verify data via the output file. The output file is the original rack ID# with the .GWT extension.
The software has overwritten the 5 tubes specified by the reweigh proces
6. For subsequent racks, begin at Step #4 again. The Tray Setup procedure must be repeated
for each rack needing specific tubes reweighed.