Software overview - 11.25 Prepare grids workflow - clean dispenser

Pressing Finish Session on the Load More Grids form begins the dispenser cleaning process. Steps from this point on are not time critical. 

IMPORTANT: Whenever you are leaving chameleon inactive for an hour or more, you must first clean the dispenser. 

The dispenser will be cleaned automatically using the water and chamclean ports in the wash station. This will take approximately eight minutes.  

fig:

The cryogen drawer can be opened to top up LN2 during the cleaning cycle by pressing the Snowflake icon in the toolbar.  

Complete removal of the cryogen bowl is not possible at this time.  

Removal of grid storage buttons is not advisable at this time.

Next article - Software overview - 11.26 Prepare grids workflow - Test dispenser